Symmetric Dimethylarginine (SDMA), Elisa kit

https://piadiagnostika.com/web/image/product.template/4082/image_1920?unique=e5e280e
Determinación cuantitativa de dimetilarginina simétrica endógena (SDMA) en suero o plasma EDTA de perros y gatos.

Referencia: EA203/96
Contenido: Unidades
Unidad de venta: Kit

569,00 € 569.0 EUR 569,00 € Impuestos no incluidos

No está disponible para venta

Esta combinación no existe.

Términos y condiciones
Shipping: 2-3 Business Days

Este kit ELISA contiene reactivos para la determinación cuantitativa de SDMA derivatizado en plasma y suero de perros o gatos.

Especificaciones

  • Muestra: plasma o suero de perro o gato

  • Contenido: 96 pocillos (8/12)

  • Rango de ensayo: perros 0.30 – 0.65 µmol/l (6.0 – 13 µg/dl), gatos: 0.30 – 0.75 µmol/l (6.0 – 15 µg/dl)

  • Sensibilidad: 0,03 µmol/l (0,6 µg/dl)

  • Tipo de test: inmunoensayo enzimático competitivo.

  • Almacenamiento a 2–8 °C

EA203/96 SDMA Kit English

Instructions for Use

SDMA vet ELISA

Enzymimmunoassay for the Quantitative Determination of Endogenem symmetrischen Dimethyl-Arginin (SDMA) in Serum or Plasma of Dogs and Cats

For Veterinary Diagnostic Only

REF EA203/96
12 * 8
2 - 8 °C

sdmavet-e_15.docx 2025-01-10

Table of Contents

Hazard Pictograms

Danger
Warning

1. Introduction and Principle of the Test

Symmetric dimethylarginine (SDMA) is a methylated arginine amino acid. SDMA is derived from intranuclear methylation of L-arginine residues and is released into the cytoplasm after proteolysis. SDMA is excreted by the kidneys.

Several studies have found that 1 in 3 cats and 1 in 10 dogs are likely to develop a kidney disease during their lifetime.

SDMA is an early biomarker of kidney function. It correlates well with glomerular filtration rate (GFR). On average, SDMA increases in chronic kidney disease (CKD) with 30 to 40% loss of kidney function. Creatinine, however, does not increase until 75% of kidney function is lost. SDMA will enable veterinarians to diagnose chronic kidney disease (CKD) much earlier than Creatinine or Cystatin C tests.

SDMA is specific for kidney function. It is not impacted by other diseases such as liver disease, cardiovascular disease, inflammatory disease and endocrine diseases. Another exciting feature of SDMA is that it is not impacted by muscle mass either, which simplifies diagnosing and monitoring CKD in thin geriatric animals, especially cats and animals with other diseases that cause muscle wasting.

The competitive SDMA-ELISA uses the microtiter plate format. SDMA is bound to the solid phase of the microtiter plate. SDMA in the samples is acylated and competes with solid phase bound SDMA for a fixed number of rabbit anti- SDMA antiserum binding sites. When the system is in equilibrium, free antigen and free antigen-antiserum complexes are removed by washing. The antibody bound to the solid phase SDMA is detected by anti-rabbit / peroxidase. The substrate TMB / peroxidase reaction is monitored at 450 nm. The amount of antibody bound to the solid phase SDMA is inversely proportional to the SDMA concentration of the sample.

2. Precautions

  • For veterinary diagnostic only. For professional use only.
  • Before carrying out the test, the valid instructions for use, as included in this kit, should be read completely and the content understood.
  • Material of animal origin used in the preparation of the kit has been obtained from animals certified as healthy but these materials should be handled as potentially infectious.
  • Individual components of different lots and test kits should not be interchanged. The expiry dates and storage conditions stated on the packaging and the labels of the individual components must be observed.
  • When handling the reagents, controls and samples, the current laboratory safety guidelines and good laboratory practice should be observed.
  • Wear protective clothing, disposable gloves, and safety goggles while performing the test.
  • Some of the components of this test kit contain hazardous substances. These components bear the appropriate hazard symbol on their label. Further information can be found in 4. Contents of the kit and on the relevant safety data sheets.
  • Avoid any actions that could result in ingestion, inhalation or injection of the reagents. Never pipette by mouth.
  • Avoid contact with individual reagents.
  • Dispose of waste according to state and local environmental protection regulations.
  • Broken glass can cause injury. Be cautious with glass vials.

3. Storage and Stability

The kit is shipped at ambient temperature. Upon arrival, store the kit at 2 - 8 °C to keep it stable until its expiry date. Once opened the kit is stable until its expiry date. The shelf life of the ready-to-use reagents is indicated on the respective bottle label. For stability of prepared reagents refer to 6.

Reagents must equilibrate to room temperature before use and be refrigerated immediately after use.

4. Contents of the Kit

MT-Strips STRIPS 12 strips

8 wells each, break apart, precoated with SDMA

Standards (1 - 6) CAL 1 - 6 6 vials

each 4 ml, ready for use

Standard 1 2 3 4 5 6
µmol/l 0 0,2 0,4 0,7 1,2 3
ng/ml 0 40 81 141 242 606
Control 1 & 2 CON 1 & 2 2 vials

each 4 ml ready for use, Range: see QC certificate

Acylation Reagent ACYL-REAG 3 vials
lyoph., dissolve contents in 3 ml Solvent before use
Acylation Buffer ACYL-BUFF 1 vial
3.5 ml, ready for use, blue coloured
Warning
Solvent SOLVENT 1 Flasche
10 ml ready for use, contains DMSO, Please note that Solvent reacts with many plastic materials including plastic trays; Solvent does not react with normal pipette tips and with glass devices
Warning
Danger
Antiserum AS 1 vial
7 ml, ready for use, Rabbit-anti-N-acyl-SDMA, yellow coloured
Warning
Enzyme Conjugate CONJ 1 vial
13 ml, ready for use, Goat-anti-rabbit-IgG-peroxidase
Warning
Wash Buffer WASH 1 vial
20 ml, conc. (50x), Dilute with dist. Water to 1000 ml total volume
Substrate SUB 1 vial
13 ml TMB Solution, ready for use
Stop Solution STOP 1 vial
13 ml, ready for use, contains 0.3M sulphuric acid, not corrosive
Reaction Plate ACYL-PLATE 1 vial
For acylation
Equalizing Reagent EQUA-REAG 1 vial
lyoph., dissolve contents with 21 ml dist. water, dissolve carefully to minimize foam formation
Foil FOIL 2 pieces
Ready for use

Additional materials and equipment required but not provided:

  • Pipettes (20, 50, 100 and 200 μl)
  • Multipette
  • Orbital shaker
  • Microplate washing device
  • Microplate photometer (450 nm)
  • Vortex mixer, roll mixer
  • Paper towels, pipette tips, timer

5. Sample Collection

The test can be performed with serum as well as with EDTA plasma. Hemolytic and lipemic samples should not be used. The samples can be stored up to 6 hours at 2 – 8 °C. For a longer storage (up to 18 months) the samples must be kept frozen at -20 °C Repeated freezing and thawing should be avoided.

6. Preparation of Reagents and Samples

6.1 Microtiter strips

Before opening the packet of strip wells STRIPS , allow it to stand at room temperature for at least 10 minutes. After opening, place any unused wells in the original foil packet with the desiccant provided. Reseal carefully and store at 2 – 8 °C.

6.2 Wash Buffer

Dilute the contents of WASH with dist. water to a total volume of 1000 ml, mix shortly. The diluted wash buffer must be stored at 2 – 8 C and is stable for 4 weeks. Should the kit be used in several runs, then prepare only the required amount of wash buffer for each run.

6.3 Equalizing Reagent

Dissolve the contents of EQUA-REAG with 21 ml dist. water, mix shortly and leave on a roll mixer for 20 minutes. Avoid excess formation of foam. The reconstituted Equalizing Reagent should be stored frozen at -20 °C and is stable until expiry date.

6.4 Acylation Reagent

Dissolve the contents of one bottle ACYL-REAG with 3 ml Solvent SOLVENT and shake for 10 minutes on an orbital shaker. After use the reagent has to be discarded. The Acylation Reagent has always to be prepared immediately before use and is stable for minimum 3 hours. The two other bottles allow a second and third run of the test. If the whole kit is to be used in one run, it is recommended to pool the dissolved contents of the two vials of Acylation Reagent. Please note that Solvent reacts with many plastic materials including plastic trays which are used as reservoir for multichannel pipettes. Solvent does not react with normal pipette tips and with glass devices. It is recommended to use a multipette, fill it directly from the vial and add the Acylation Reagent to the wells.

All other reagents are ready for use.

6.5 Preparation of Samples (Acylation)

Bring all reagents to room temperature and mix them carefully, avoid development of foam. The wells of the reaction plate ACYL-PLATE for the acylation can be used only once. Please mark the respective wells before use to avoid repeated use.

  • Pipette each 20 μl standard 1 - 6 CAL 1–6, each 20 μl control 1 & 2 CON 1 & 2 and each 20 μl patient sample into the respective wells of the Reaction Plate ACYL-PLATE (duplicates are recommended).
  • 2. Pipette 20 μl Acylation Buffer ACYL-BUFF into each well.
  • 3. Pipette 200 μl reconstituted Equalizing Reagent EQUA-REAG (see 6.3) into each well.
  • 4. Mix the reaction plate for 10 seconds on an orbital shaker.
  • 5. Pipette 50 μl of freshly prepared Acylation Reagent ACYL-REAG (see 6.4) into each well, mix immediately. It is recommended to use a multipette, fill it directly from the vial and add the Acylation Reagent to each well. Colour changes to violet.
  • 6. Incubate for 20 minutes at room temperature (approx. 20 °C) on an orbital shaker. Do not cover wells or plate, leave the plate open on the shaker.

Take each 20 μl of the acylated samples for the SDMA-ELISA.

7. Test Procedure ELISA

Bring all reagents to room temperature and mix them carefully, avoid development of foam.

7.1 Sample Incubation

Transfer each 20 μl acylated Standards 1 to 6, 20 μl acylated controls and 20 μl acylated samples from the Reaction Plate into the respective wells of the coated microtiter strips STRIPS . Pipette 50 μl Antiserum AS into each well. Cover the plate with adhesive foil FOIL and incubate Microtiter Strips for 90 minutes at room temperature (20 – 25 °C) on an orbital shaker

7.2 Washing

Discard or aspirate the contents of the wells and wash thoroughly with each 300 μl diluted Wash Buffer WASH (see 6.2). Repeat the washing procedure 4 times. Remove residual liquid by tapping the inverted plate on clean absorbent paper.

7.3 Conjugate Incubation

Pipette 100 μl enzyme conjugate CONJ into each well. Incubate for 30 minutes at room temperature on an orbital shaker.

7.4 Washing

Repeat step 7.2.

7.5 Substrate Incubation

Pipette 100 μl Substrate SUB into each well and incubate for 25 ± 5 minutes at room temperature on an orbital shaker.

7.6 Stopping

Pipette 100 μl Stop Solution STOP into each well and mix on an orbital shaker for approx. 30 seconds.

7.7 Reading

Within 15 minutes, read the optical density at 450 nm (reference wavelength between 570 and 650 nm) in a microplate photometer.

8. Calculation of the Results

Standard 1 2 3 4 5 6
µmol/l 0 0,2 0,4 0,7 1,2 3
ng/ml 0 40 81 141 242 606

Conversion factor: 1 μmol/l = 202 ng/ml = 20.2 μg/dl

The concentration of the standards (x-axis, logarithmic) are plotted against their corresponding optical density (y-axis, linear). Cubic spline, 4 parameter or similar iteration procedures are recommended for evaluation of the standard curve.

The concentration of the controls and samples can be read directly from this standard curve by using their average optical density.

Typical standard curve:

SDMA Elisa calculation

Quality Control: Test results are valid only if the kit controls are within the ranges specified on the QC Certificate. Otherwise, the test should be repeated.

9. Assay Characteristics

9.1 Expected Values (Serum, EDTA-Plasma)

Dogs: 0,30 – 0,65 µmol/l (6,0 – 13 µg/dl)

Cats: 0,30 – 0,75 µmol/l (6,0 – 15 µg/dl)

The reference ranges given above should only be taken as a guideline. It is recommended that each laboratory should establish its own reference values.

9.2 Sensitivity

0,03 µmol/l (0,6 µg/dl)

9.3 Recovery

9.3.1 Recovery Cat

Increasing amounts of SDMA were added to a cat serum sample. Each spiked sample was assayed. The analytical recovery of SDMA was estimated at 10 different concentrations by using the theoretically expected and the actually measured values. The mean recovery from all concentrations was 101%.

added
[µmol/l]
measured
[µmol/l]
expected
[µmol/l]
% recovery
0,00 0,58
0,12 0,68 0,70 97
0,24 0,77 0,82 91
0,35 0,86 0,93 92
0,45 0,91 1,03 88
0,55 0,99 1,13 88
0,65 1,24 1,23 101
0,77 1,71 1,35 127
1,04 1,90 1,62 117
1,35 1,95 1,93 101
1,65 2,29 2,23 103
mean value 101

9.3.2 Recovery Dog

Increasing amounts of SDMA were added to a dog serum sample. Each spiked sample was assayed. The analytical recovery of SDMA was estimated at 10 different concentrations by using the theoretically expected and the actually measured values. The mean recovery from all concentrations was 104 %.

added [µmol/l] measured [µmol/l] expected [µmol/l] % recovery
0,00 0,54
0,12 0,74 0,66 112
0,24 0,74 0,78 95
0,35 0,86 0,89 97
0,45 0,94 0,99 95
0,55 1,01 1,09 93
0,65 1,19 1,18 101
0,77 1,51 1,31 115
1,04 1,73 1,58 109
1,35 2,19 1,89 116
1,65 2,23 2,19 102
mean value 104

9.4 Linearity

9.4.1 Linearity Cat

The linearity of the ELISA method was investigated using seven different dilutions of a cat serum sample. The mean linearity from all dilutions was 96%.

dilution measured [µmol/l] recalculated value [µmol/l] recovery %
orig. 2,09
4 + 1 1,73 2,16 103
2 + 1 1,30 1,95 93
1 + 1 0,95 1,90 91
1 + 2 0,59 1,77 85
1 + 3 0,52 2,08 100
1 + 5 0,33 1,98 95
1 + 7 0,28 2,24 107
mean recovery 96

9.4.2 Linearity Dog

The linearity of the ELISA method was investigated using seven different dilutions of a dog serum sample. The mean linearity from all dilutions was 92%.

dilution measured [µmol/l] recalculated value [µmol/l] recovery %
orig. 1,88
4 + 1 1,30 1,63 87
2 + 1 1,21 1,82 97
1 + 1 0,92 1,84 98
1 + 2 0,59 1,77 94
1 + 3 0,43 1,72 91
1 + 5 0,28 1,68 89
1 + 7 0,20 1,60 85
mean recovery 92

9.5 Specificity (Cross Reactivity)

Structural related components were tested for possible interference with the antisera against SDMA used in the ELISA method. The tested compounds were Arginine, Homoarginine, Monomethylarginine (NMMA) and ADMA.

Substance ED-50-Value [µmol/l] Cross Reactivity (%)
SDMA 1,39 100
ADMA 84 1,2
NMMA 182 0,76
Homoarginine 2807 0,05
Arginine 8574 0,016

9.6 Reproducibility

The reproducibility of the ELISA method was investigated by determining the intra- and inter-assay-coefficient of variation (cv) by repeated measurements of different serum samples with different SDMA concentrations (concentrations in μmol/l):

9.6.1 Intra-Assay Cat

Sample n = mean value SD cv (%)
K1 40 0,837 0,064 7,6
K2 40 0,862 0,049 5,7
K3 40 0,770 0,058 7,7

9.6.2 Intra-Assay Dog

Sample n = mean value SD cv (%)
H1 40 0,531 0,061 11,5
H2 40 0,804 0,068 8,5
H3 40 0,776 0,046 5,9

9.6.3 Inter-Assay Cat

Sample n = mean value SD cv (%)
K1 32 0,49 0,037 7,5
K2 32 0,59 0,043 7,3
K3 32 0,72 0,079 11,0
K4 32 0,74 0,058 7,8
K5 32 0,84 0,068 8,1
K6 32 0,84 0,982 9,7

9.6.4 Inter-Assay Dog

Sample n = mean value SD cv (%)
H1 32 0,49 0,042 8,5
H2 32 0,66 0,046 7,0
H3 32 0,76 0,063 8,2
H4 32 0,93 0,069 7,4
H5 32 1,21 0,128 10,6
H6 32 1,35 0,152 11,3

9.7 Correlation ELISA to LC-MS/MS

Correlation ELISA to the LC-MS/MS method

SDMA Dog

Correlation ELISA to the LC-MS/MS method in dogs

SDMA cat

Correlation ELISA to the LC-MS/MS method in cats

10. Literature

  • Josipa Kuleš, Petra Bilić, Blanka Beer Ljubić,Jelena Gotić, Martina Crnogaj, Mirna Brkljačić, Vladimir Mrljak. Glomerular and tubular kidney damage markers in canine babesiosis caused by Babesia canis Ticks and Tick-borne Diseases (2018) 9 1508 - 1517
  • Bode-Böger S.M., Scalera F., Kielstein J.T., Martens-Lobenhoffer J., Breithardt G., Fobker M., Reinecke H. Symmetrical Dimethylarginine: A new combined parameter for renal function and extent of coronary artery disease J. Am. Soc. Nephrol. (2006) 17: 1128-1134
  • Kielstein J.T., Salpeter S.R., Bode-Böger S.M., Cooke J.P., Fliser D. Symmetric dimethylarginine (SDMA) as endogenous marker of renal function – a meta-analysis Nephrol. Dial. Transplant (2006) 21: 2446 – 2451

11. Changes to declare

Version _15: Changes are highlighted in gray.

Version _14: Changes in section 4 are highlighted in gray.

Version _13: Additions/Changes are highlighted in gray.

Version _12: Hazard symbol “Warning” was removed from POD Conjugate in section 4, as no longer required. Grey highlighting as in version _11 has been removed.

Version _11: IFU has been re-formatted. Precautions and Calculation of the Results have been complemented (highlighted in grey). Component names as printed on labels were included in pipetting schemes to provide greater clarity. No changes have been made to components or execution of protocols.

Pipetting Scheme - Sample Preparation

Standards Controls Samples
ACYL-PLATE:
CAL 1–6 µl 20
CON 1 & 2 µl 20
Sample µl 20
ACYL-BUFF µl 20 20 20
recon. EQUA-REAG µl 200 200 200

Shake for 10 seconds

recon. ACYL-REAG, fresh µl 50 50 50

Incubate for 20 minutes at room temperature on an orbital shaker Take 20 μl each for ELISA

Pipetting Scheme - ELISA

Acyl. Acyl. Acyl.
Acyl. Standards Acyl. Controls Acyl. Samples
STRIPS:
Transfer from ACYL-PLATE into STRIPS µl 20 20 20
AS µl 50 50 50

Cover frame with FOIL and incubate on an orbital shaker

for 90 minutes at room temperature

Wash 4 x with 300 μl WASH per well

CONJ µl 100 100 100

Incubate for 30 minutes at room temperature on an orbital shaker

Wash 4 x with 300 μl WASH per well

SUB µl 100 100 100

Incubate for 25 ± 5 minutes at room temperature on an orbital shaker

SUB µl 100 100 100

Shake for 30 seconds

Within 15 minutes, read absorbance at 450 nm (ref. 570 – 650 nm)